FLUORESCENCE
MICROSCOPE
A fluorescence microscope is basically a conventional light microscope with added features and components that extend its capabilities.
A fluorescence microscope is basically a conventional light microscope with added features and components that extend its capabilities.
·
A conventional microscope uses light
to illuminate the sample and produce a magnified image of the sample.
·
A fluorescence microscope uses a much
higher intensity light to illuminate the sample. This light excites
fluorescence species in the sample, which then emit light of a longer
wavelength. A fluorescent microscope also produces a magnified image of the
sample, but the image is based on the second light source -- the light
emanating from the fluorescent species -- rather than from the light originally
used to illuminate, and excite, the sample.
Working
Principle:
In most
cases the sample of interest is labelled with a fluorescent substance known as
a fluorophore and then illuminated through the lens with the higher energy
source. The illumination light is absorbed by the fluorophores (now attached to
the sample) and causes them to emit a longer lower energy wavelength light.
This fluorescent light can be separated from the surrounding radiation with
filters designed for that specific wavelength allowing the viewer to see only
that which is fluorescing.
The basic
task of the fluorescence microscope is to let excitation light radiate the
specimen and then sort out the much weaker emitted light from the image. First,
the microscope has a filter that only lets through radiation with the specific
wavelength that matches your fluorescing material. The radiation collides with
the atoms in your specimen and electrons are excited to a higher energy level.
When they relax to a lower level, they emit light. To become detectable
(visible to the human eye) the fluorescence emitted from the sample is
separated from the much brighter excitation light in a second filter. This
works because the emitted light is of lower energy and has a longer wavelength
than the light that is used for illumination.
Most of the
fluorescence microscopes used in biology today are epi-fluorescence
microscopes, meaning that both the excitation and the observation of the
fluorescence occur above the sample. Most use a Xenon or Mercury arc-discharge
lamp for the more intense light source.
Instrumentation:
Nearly all fluorescence microscopes have
following basic parts below.
·
Focus
Illumination
·
Collector
for emitted fluorescence
·
Dichroic
mirror
·
Filter:
1.
Focus the
illumination (excitation) light on the sample.
In order to excite
fluorescent species in a sample, the optics of a fluorescent microscope must
focus the illumination (excitation) light on the sample to a greater extent
than is achieved using the simple condenser lens system found in the illumination
light path of a conventional microscope.
2. Collect the emitted fluorescence:
This type of
excitation-emission configuration, in which both the excitation and emission
light travel through the objective, is called epifluorescence. The key to the optics
in an epifluorescence microscope is the separation of the illumination
(excitation) light from the fluorescence emission emanating from the sample. In
order to obtain either an image of the emission without excessive background
illumination, or a measurement of the fluorescence emission without background
"noise", the optical elements used to separate these two light
components must be very efficient.
In
a fluorescence microscope, a dichroic mirror is used to separate the
excitation and emission light paths. Within the objective, the excitation
emission share the same optics.
The excitation light reflects off
the surface of the dichroic mirror into the objective.
The fluorescence emission passes
through the dichroic to the eyepiece or detection system.
The dichroic mirror's
special reflective properties allow it to separate the two light paths. Each
dichroic mirror has a set wavelength value -- called the transition
wavelength value -- which is the wavelength of 50% transmission. The mirror
reflects wavelengths of light below the transition wavelength value and
transmits wavelengths above this value. This property accounts for the name
given to this mirror (dichroic, two color). Ideally, the wavelength of the
dichroic mirror is chosen to be between the wavelengths used for excitation and
emission.
The dichroic mirror is
a key element of the fluorescence microscope, but it is not able to perform all
of the required optical functions on its own. Typically, about 90% of the light
at wavelengths below the transition wavelength value are reflected and about
90% of the light at wavelengths above this value are transmitted by the
dichroic mirror. When the excitation light illuminates the sample, a small
amount of excitation light is reflected off the optical elements within the
objective and some excitation light is scattered back into the objective by the
sample. Some of this "excitation" light is transmitted through the
dichroic mirror along with the longer wavelength light emitted by the sample.
This "contaminating" light would otherwise reach the detection system
if it were not for another wavelength selective element in the fluorescence
microscope: an emission filter.
Figure 1: Optical diagram of
Fluorescent Microscope
Two filters are used along with the
dichroic mirror:
Excitation
filter -- In order to
select the excitation wavelength, an excitation filter is placed in the
excitation path just prior to the dichroic mirror.
Emission
filter -- In order to more
specifically select the emission wavelength of the light emitted from the
sample and to remove traces of excitation light, an emission filter is placed
beneath the dichroic mirror. In this position, the filter functions to both
select the emission wavelength and to eliminate any trace of the wavelengths
used for excitation.
These
filters are usually a special type of filter referred to as an interference
filter, because of the way in which it blocks the out of band transmission.
Interference filters exhibit an extremely low transmission outside of their
characteristic bandpass. Thus, they are very efficient in selecting the desired
excitation and emission wavelengths.
Applications:
The refinement of epi-fluorescent microscopes
and advent of more powerful focused light sources, such as lasers, has led to
more technically advanced scopes such as the confocal laser scanning
microscopes and total internal reflection fluorescence microscopes (TIRF).
CLSM's are invaluable tools for producing
high resolution 3-D images of sub-surfaces in specimens such as microbes. Their
advantage is that they are able to produce sharp images of thick samples at
various depths by taking images point by point and reconstructing them with a
computer rather than viewing whole images through an eyepiece.
These microscopes are often used for -
·
Imaging structural components of small
specimens, such as cells
·
Conducting viability studies on cell
populations (are they alive or dead?)
·
Imaging the genetic material within a cell
(DNA and RNA)
·
Viewing specific cells within a larger
population with techniques such as FISH
0 comments:
Post a Comment