NANODROP
SPECTROPHOTOMETRY
(Quantification of Nucleic Acids)
(Quantification of Nucleic Acids)
NanoDrop is
a Spectrophotometer that measure 0.5- 2 ml samples
with high accuracy and reproducibility. The NanoDrop model offers the convenience
of both the NanoDrop patented sample retention technology and a traditional
cuvette for sample measurements.
Principle: It operates on Beer’s law.
When monochromatic light (light of
a specific wavelength) passes through a solution there is usually a
quantitative relationship (Beer's law) between the solute concentration and the
intensity of the transmitted light, that is, the more concentrated the specimen
is, the less light is transmitted through it.
The sample retention system in
NanoDrop Spectrophotometry employs surface tension to hold the sample in place
between two optical fibers. This enables the measurement of very highly
concentrated samples without the need for dilutions. Using this technology, the full spectrum (190
– 840 nm) NanoDrop Spectrophotometers have the capability to measure sample
concentration up to 200 times more concentrated than samples measured using the
standard cuvette.
Sample Retention Pedestal Measurements:
A 1 - 2 µL
sample is pipetted onto a measurement pedestal.
A smaller,
0.5 µL volume sample, may be used for concentrated nucleic acid and protein
A280 samples.
A fiber
optic cable (the receiving fiber) is embedded within this pedestal.
A second
fiber optic cable (the source fiber) is then brought into contact with the
liquid sample causing the liquid to bridge the gap between the ends of the two
fibers.
A pulsed
xenon flash lamp provides the light source and a spectrometer utilizing a
linear CCD array analyzes the light passing through the sample.
The instrument
is controlled by PC based software, and the data is stored in workbook files
(*.twbk) on the PC.
Pedestal Sample Size Requirements: Although
sample size is not critical, it is essential that a liquid column is formed
when using the pedestal option so that the pathlength between the upper and
lower measurement pedestals is bridged with sample. The dominant factor determining the surface
tension of a droplet is the hydrogen bonding of the lattice of water molecules
in solution. Generally, all additives (including protein, DNA, RNA, buffer
salts and detergent-like molecules) can reduce the surface tension by
interfering with the hydrogen bonding between water molecules. Although 1 µL
volumes are usually sufficient for most sample measurements, increasing the
sample size to 2 µL will ensure proper column formation for samples with
reduced surface tension.
Sample volumes are sufficient to ensure reproducibility
- Aqueous
solutions of nucleic acids: 1
µL
- Purified
protein: 2
µL
- Bradford,
BCA, Lowry or Protein Pierce 660 nm assays: 2 µL
- Microbial
cell suspensions: 2
µL
Note: It is best
to use a precision pipettor (0-2 µL) with precision tips to ensure that
sufficient sample (1-2 µL) is delivered. Lower precision pipettors (0-10 µL and
larger) are not as good at delivering 1 µL volumes to the measurement pedestal.
If the user is unsure about the sample
characteristics or pipettor accuracy, a 2 µL sample volume is recommended.
Pedestal Basic Use
- Raise
the sampling arm and pipette the sample onto the lower measurement
pedestal.
- Lower the sampling arm and initiate a spectral
measurement using the software on the PC. The sample column is
automatically drawn between the upper and lower pedestals and the
measurement is made.
- When the measurement is complete, raise the
sampling arm and wipe the sample from both the upper and lower pedestals
using a dry, lint-free laboratory wipe. Simple wiping prevents sample
carryover in subsequent measurements for samples varying by more than 1000-fold
in concentration.
Cuvette Measurements
The NanoDrop will accept 10 mm
cuvettes up to 48 mm tall. When
measuring samples using micro, semi-micro, or ultra-micro cuvettes, we
recommend using masked cuvettes. Masked
cuvettes ensure that all light hitting the detector has passed through the
sample. Unmasked cuvettes can allow light to hit the detector that has not
passed through the sample. Expected
differences between unmasked plastic cuvettes can lead to significant
measurement error, especially at low concentrations.
Cuvette Sample Size Requirements:
It is essential to ensure that the
sample volume in the cuvette is adequate to allow the light to pass through a
representative portion of the sample when making a measurement.
Cuvette Basic Use
- Add the sample to the cuvette and ensure that the
volume is sufficient to cover the light path.
- Raise the sampling arm and insert the cuvette
into the instrument. Insert the cuvette noting the direction of the light
path indicated by the etched arrow.
- The sampling arm must be lowered into the down
position for cuvette measurements.
- Initiate a spectral measurement using the
software on the PC.
- When the measurement is complete, remove the
cuvette, rinse thoroughly and dry between samples.
Requirements:
- Sample to be measured
- NanoDrop
- Micropipette with tips
- Lint free lab wipes
- Purified water
- Blanking solution (D/W, TE, TB Tris or others depending on your
sample)
Applications:
DNA Analysis (After extraction): Nucleic acid (DNA/RNA)
Qualitative analysis Quantitative
analysis
Size-Quality Purity-
Yield
Agarose gel electrophoresis Spectrophotometry
Why Quantify?
• To check
concentration and purity of DNA/RNA present in the solution.
• Determine
if samples are useful for downstream applications like: PCR, Restriction
digests etc.
Advantages of the nanodrop
• It is a
simple machine and economical on space.
• Easy-to-use
spectrophotometer
• Can measure
small volumes of DNA, RNA and protein concentrations.
NanoDrop Spectrophotometric measures
Measures DNA, RNA (A260) and Proteins (A280) concentrations
and sample purity (260:280).
Absorbance at 260 nm
• Nucleic
acids absorb UV light at 260 nm due to the aromatic base moieties within their
structure. Purines (thymine, cytosine and uracil) and pyrimidines (adenine and
guanine) both have peak absorbances at 260 nm, thus making it the standard for
quantitating nucleic acid samples.
Absorbance at 280 nm
• The 280 nm
absorbance is measured where proteins and phenolic compounds have a strong
absorbance. Similarly, the aromaticity of phenol groups of organic compounds
absorbs strongly near 280 nm.
Absorbance at 230 nm
• Many
organic compounds have strong absorbances at around 225 nm. In addition to
phenol, TRIzol, and chaotropic salts, the peptide bonds in proteins absorb
light between 200 and 230 nm.
NanoDrop Spectrophometric measurements
• The A260/280
ratio shows the purity of the sample analyzed.
Nucleic Acid Type
|
Approximate A260/A280 Ratio
|
Pure DNA
|
1.8
|
Pure RNA
|
2.0
|
Pure Protein
|
0.57
|
• A260/230 ratio
indicates the presence of organic contaminants, such as (but not limited to):
phenol, TRIzol, chaotropic salts and other aromatic compounds.
• Samples
with 260/230 ratios below 1.8 are considered to have a significant
amount of these contaminants
• Various
nanodrop models are available e.g Nanodrop 8000, Nanodrop lite and Nanodrop
2000/2000c which is currently available at BecA labs.
Note that: Because of
the low concentrations, sometimes it is difficult to assess the purity of the
samples (esp. RNA) by analyzing the A260/280 and A260/230 ratios.
Quantification cannot be assessed by the NanoDrop because they are outside the
lowest concentrations the NanoDrop is designed to measure. A more sensitive method such as an Agilent
Bioanalyzer or Qubit analysis is recommended.